The Agilent 1200 Series G1367B High Performance Autosampler comes refurbished and adds maximum flexibility and fast injection cycles, whenever high sample throughput and speed of analysis are required. The standard metering device (for the G1367B) provides injection volumes from 0.1°C 100 µl. A multi-draw kit extends the range up to 1500 µl.
Specifications:
- GLP Features
- Early Maintenance Feedback (EMF)
- Electronic records of maintenance and errors
- Communications
- Controller-area network (CAN)
- RS232C
- APG-remote standard
- optional four external contact closures and BCD vial number output
- Safety Features
- Leak detections and safe leak handling
- Low voltages in maintenance areas
- Error detection and display
- Injection Range
- 0.1 – 100 µl in 0.1 µl increments Up to 1500 µl with multiple draw (hardware modification required)
- Precision
- Typically < 0.25% RSD from 5 – 100 µl
- Typically < 1% RSD from 1 – 5 µl variable volume
- Pressure Range
- 400 bar (5880 psi)
- Sample Viscosity Range
- 0.2 – 5 cp
- Sample Capacity
- 2 × well plates (MTP) + 10 × 2 ml vials 108 x 2-mL vials in2 x 54 vial plate plus 10 additional 2 mL vials 30 x 6-mL vials in 2 x 15 vial plate plus 10 additional 2 mL vials
- 54 Eppendorf tubes (0.5/1.5/2.0mL) in 2 x 27
- Eppendorf tube plate
- Also compatible with the Agilent 1200 Series sample capacity extension for further expansion of the sample capacity
- Injection Cycle Time
- Typically < 30 s using following standard conditions:
- Default draw speed: 200 µl/min
- Default eject speed: 200 µl/min
- Injection volume: 5 µl
- Typically < 30 s using following standard conditions:
- Carry-Over
- Typically < 0.01% using the following conditions:
- Column: 125 x 4 mm Hypersil ODS, 5 µm
- Mobile phase: Water/Acetonitrile = 80/20
- Flow rate: 1 ml/min
- Injection volume: 1 µl caffeine (1 mg/ml) 5 µl water to test carryover
- Outside wash of needle before injection: 20 sec with water using flush port
- Typically < 0.01% using the following conditions: