This is a refurbished Agilent 1200 Series G1312A Binary Pump.
The Agilent 1200 G1312A Binary Pump is a high-performance, binary pumping system for use in liquid chromatography. It features a compact design and operates using a binary pumping mechanism, which allows for high pressure operation and efficient sample delivery. The pump features a high-resolution, low-pulsation flow rate, making it ideal for use in demanding applications.
Type | Specification |
Hydraulic system | Two dual piston in series pumps with proprietary servo-controlled variable stroke drive, floating piston design and active inlet valve |
Setable flow range | Set points 0.001 – 5 ml/min, in 0.001 ml/min increments |
Flow range | 0.1 – 5.0 ml/min |
Flow precision | <0.07% RSD, or <0.02 min SD whatever is greater, based on retention time at constant room temperature |
Flow accuracy | +1% of 10µl/min whatever is greater |
Pressure | Operating range 0 400 bar (0-5880 psi) up to 5 ml/min |
Pressure pulsation | < 2% amplitude (typically < 1%), at 1 ml/min isopropanol, at all pressures > 1 MPa |
Compressibility compensation | User-selectable, based on mobile phase compressibility |
Recommended pH range | 1.0 – 12.5, solvents with pH <2.3 should not contain acids which attack stainless steel |
Gradient formation | High-pressure binary mixing, delay volume 180-480 µl without mixer, 600 – 900 µl with mixer, dependent on back pressure |
Composition range | 1 – 99 % or 5 µl/min per channel, whatever is greater |
Composition precision | <0.5% absolute |
Composition accuracy | 0.15% RSD, at 1 ml/min |
Control and data evaluation | Agilent Chem Station for LC |
Analog output | For pressure monitoring, 2mV/bar, one output |
Communications | Controller-area network (CAN), GPIB, RS-232C, APG Remote: ready, start, stop and shut-down signals, LAN optional |
Safety and maintenance | Extensive diagnostics, error detection and display (through handheld controllers G4208 A, G1323 B and Agilent Chem Station), leak detection, safe leak handling, leak output signal for shut down of pumping system. Low voltage in major maintenance areas |
GLP features | Early maintenance feedback (EMF) for continuous tracking of instruments usage in terms of seal wear and volume of pumped mobile phase with user-settable limits and feedback messages. Electronic records of maintenance and errors. |
Housing | All materials recyclable |